Analyzer FAQs

Frequently Asked Questions

What can I do if the moisture reading on my analyzer is higher than I think it should be?

DILO can recommend some steps to take if you suspect the moisture reading to be high:

  • Ensure the couplings and test hoses being utilized are wiped down of any residual moisture.
  • Ensure no rubber is being used for the testing equipment i.e. regulator diaphragm or testing hose.
  • Pressurize the inlet hose and check for leaks using a leak detector. Leaks can cause the moisture values to increase.
  • Conduct the “Purge” or “Rinse” function on the analyzer and then run additional tests.
  • If you have a 3-038R, try turning off automatic pump back and running additional tests.
  • Lastly, regulate and connect a bottle of Nitrogen to the inlet of the Analyzer and run a couple of tests to dry out the internal contents of the device. Reconnect to previous gas source and run a test.

If these steps do not correct the issue, please contact us.

What can I do if my analyzer purity reading is lower than I think it should be?

DILO can recommend some steps to take if you suspect the purity reading is off:

  • Pressurize the inlet hose and check for leaks using a leak detector. Leaks can cause the moisture values to increase.
  • Conduct the “Purge” or “Rinse” function on the analyzer and then run additional tests.

If these steps do not correct the issue, please contact us.

What can I do if the “Internal Storage Pressure to High” warning occurs on my analyzer?

DILO recommends that you conduct a manual pump back function. Please refer to the operating manual or Quick-Start Guide for step-by-step procedures. If these steps do not correct the issue, please .